Negative Pressure HEPA Filtration Unit for Isolation Rooms and Other High Risk Areas Introduction, Parts of Instrument Unit, Working Mechanisms, Applications and Keynotes
Table of Contents
“Negative Pressure HEPA Filtration Unit” is useful for hospital (isolation ward that keeps patients with infectious illnesses or immunocompromised individuals who are susceptible to infections from others, away from other patients, visitors, and healthcare workers) and molecular laboratory (Sample Aliquoting and RNA Extraction Room to minimize cross contamination and infection to laboratory personnel). Negative air pressure is the by which the air pressure is lower in one site in comparison to another. Regarding negative room air pressure, the air pressure inside a given room is lower than the pressure outside the room, generating air to flow into the room from the outside.
The certified HEPA air filter is highly effective in removing microbes-sized particulate and it demonstrates efficiency on 0.3-micron particles. 0.3-micron particles are representative of most difficult particle size to capture, even more difficult than particles smaller in size. This installation unit will work roughly representative of minimum efficiency for virus, bacteria and other airborne particulate removal.
The main parts of unit has cabinet, base and motor.
Cabinet: It is made of heavy gauge steel, welded cabinet with chemical resistant powder coat finish.
Base: It is standard base includes four swivel casters.
Motor: It is dynamically balanced, precision external rotor motor to insure quiet, energy efficient and reliable operation.
| Air Flow Capability | Infinitely adjustable from 115 CFM to 1000 CFM |
| Sound Levels | 35 dBA @ 115 CFM, 66 dBA @ 775 CFM at 4 feet from the unit |
| Power requirements | 120 Vac, 60 Hz, 4 Amps, 400 Watts OR • 230 Vac, 50 Hz, 2 Amps, 400 Watts |
| Prefilter | (80-85% ASHRAE efficient) |
| Primary Filter | • HEPA (≥ 99.97 DOP efficient on 0.3 microns) • HEPA filter is oversized with 131 sq. ft. of media • Oversized HEPA allows for minimum pressure drop • Oversized HEPA allows for maximum filter life and minimum handling • Each HEPA filter individually tested and certified |
| Standard Accessories Included with every XJ-2 | • Individually tested HEPA • Window Adapter Kit to exhaust filtered air • 6” diameter flex hose • Dirty Filter Gage • Variable speed control • Portable Base • Prewired with electrical cord and molded plug |
| Optional Accessories | • Security Cover • Wall Mounting Kit • Recirculation Plenum |
The working mechanism of ‘Negative Pressure HEPA Filtration Unit’ completes into following steps-
Interception: Particle comes within one particle radius of a filter fiber and then the articles remain trapped or “filtered” by the filter media due to Van der Waals forces. Larger particles are easier to capture by Interception.
Inertial Impaction or Impingement: Heavier particles collide with filter fibers due to their inertia as air makes its way through the torturous path around the filter fibers. Larger particles are easier to filter by inertial impaction.
Diffusion: Particles increase in random motion because of to Brownian motion as particle size decreases. Greater random motion increases probability of particle to make contact with filter fiber and smaller particles are actually easier to capture due to diffusion mechanism.
Straining or Sieving: It is simple and the last phase of working of ‘Negative Pressure HEPA Filtration Unit’.
Negative pressure is monitored and validated by following ways-
Frequency of negative pressure monitoring is mandatory in following situations-
CDC guidelines call for HEPA efficiency- HEPA efficiency to be ≥ 99.97% on 0.3-micron particles per military standard 282
The CDC Guidelines for Isolation Rooms are as follows-
Encircle the correct Answer of the Following question. 1) What is the full form of…
No, several crucial velocity parameters on the screen are currently operating far outside their safe…
Introduction Also known as a microcentrifuge, this compact tool is a staple in biological, clinical,…
Introduction Mycobacteriology specimen packaging and shipping involve the standardized containment and transport of biological materials…
Introduction An Ultra-Low Temperature (ULT) Freezer typically operates within the temperature range of -400C to…
Introduction An N95 respirator is a specialised respiratory protective device certified by the National Institute…
View Comments
Thanks for any other informative website. Where else may I get that kind of information written in such an ideal way? I've a venture that I am simply now working on, and I've been on the glance out for such information.
Hi! Do you know if they make any plugins to protect against hackers? I'm kinda paranoid about losing everything I've worked hard on. Any tips?
I view something truly interesting about your website so I saved to fav.
It’s hard to search out educated folks on this topic, however you sound like you already know what you’re talking about! Thanks
I may need your help. I tried many ways but couldn't solve it, but after reading your article, I think you have a way to help me. I'm looking forward for your reply. Thanks.
I really like reading through a post that can make men and women think. Also, thank you for allowing me to comment!